How often must stability samples be tested?

Stability testing & climatic zones · Testing frequency. Verbatim excerpts from each authority's document, with the section.

SFDA Saudi Food and Drug Authority

GCC Stability Guidelines v3.3 (2018) — archived copy

For products with a proposed shelf-life of at least 12 months, the frequency of testing at the long-term storage condition should normally be every three months over the first year, every six months over the second year and annually thereafter throughout the proposed shelf-life (e.g., 0, 3, 6, 9, 12, 18, 24, 36 months).GCC v3.3 (archived) · 3.2.5 Testing frequency ↗
At the accelerated storage condition, a minimum of three time points, including the initial and final time points (e.g. 0, 3 and 6 months), from a six-month study is recommended.GCC v3.3 (archived) · 3.2.5 Testing frequency ↗

EU European Union

ICH Q1A(R2) · EMA guidelines

For products with a proposed shelf life of at least 12 months, the frequency of testing at the long term storage condition should normally be every 3 months over the first year, every 6 months over the second year, and annually thereafter through the proposed shelf life.ICH Q1A(R2) · 2.2.6 Testing Frequency ↗
At the accelerated storage condition, a minimum of three time points, including the initial and final time points (e.g., 0, 3, and 6 months), from a 6-month study is recommended.ICH Q1A(R2) · 2.2.6 Testing Frequency ↗

US FDA US Food and Drug Administration

Q1A(R2) · Q1E · ANDA guidance

For products with a proposed shelf life of at least 12 months, the frequency of testing at the long-term storage condition should normally be every 3 months over the first year, every 6 months over the second year, and annually thereafter through the proposed shelf life.FDA Q1A(R2) · 2.2.6 ↗
At the accelerated storage condition, a minimum of three time points, including the initial and final time points (e.g., 0, 3, and 6 months), from a 6-month study is recommended.FDA Q1A(R2) · 2.2.6 ↗
When testing at the intermediate storage condition is called for as a result of significant change at the accelerated storage condition, a minimum of four time points, including the initial and final time points (e.g., 0, 6, 9, 12 months), from a 12-month study is recommended.FDA Q1A(R2) · 2.2.6 ↗

WHO World Health Organization

TRS 1010 Annex 10 · zone table

For products with a proposed shelf life of at least 12 months, the frequency of testing at the long-term storage condition should normally be every three months over the first year, every six months over the second year and annually thereafter throughout the proposed shelf life.WHO TRS 1010 Annex 10 · 2.2.6 Testing frequency ↗
At the accelerated storage condition, a minimum of three time points, including the initial and final time points (e.g. 0, 3 and 6 months), from a six-month study is recommended.WHO TRS 1010 Annex 10 · 2.2.6 Testing frequency ↗
When testing at the intermediate storage condition is called for as a result of significant change at the accelerated storage condition, a minimum of four time points, including the initial and final time points (e.g. 0, 6, 9 and 12 months), from a 12-month study is recommended.WHO TRS 1010 Annex 10 · 2.2.6 Testing frequency ↗