Regulatory PathRegulatory requirements compared

SFDA item 87 · IV. PMF/VAMF

Replacement or addition of blood containers (e.g. bags, bottles)

Variation category under SFDA, EU and US FDA rules for this change, with the variant wording as published.

Open in the Variation Compass

SFDA Saudi Food and Drug Authority

Guidelines for Variation Requirements v6.3 (DS-G-006), item 87 — official PDF, p. 77 ↗

VariantChangeType
aThe new blood containers had MDMA.Type IA
bThe new blood containers do not have MDMAType II

Conditions to be fulfilled and documentation to be supplied: see the official guideline, item 87.

EU European Union

Commission Guidelines C/2025/5045 — EUR-Lex ↗

M.12 · Addition or replacement of blood containers (e.g. bags, bottles)

VariantChangeType
aThe new blood containers are CE-markedType IA
bThe new blood containers are not CE-marked and there is no impact on the quality criteria of the blood in the containerType IB
cThe new blood containers are not CE-marked and there is potentially an impact on the quality criteria of the blood in the containerType II

Note: CE-marked containers are M.12(a); containers that are not CE-marked are M.12(b) or (c), depending on the impact on blood quality.

US FDA US Food and Drug Administration

21 CFR 314.70 and FDA guidance — Outside 21 CFR 314.70 — CBER-regulated containers; 21 CFR 601.12

ChangeCategoryReference
Container change is a collection-establishment license matterNo US equivalent

Note: Blood and plasma containers are FDA (CBER)-regulated products in their own right, and switching them is a collection-establishment license matter (21 CFR 601.12), not a PMF or NDA/ANDA variation.

Other items in IV. PMF/VAMF

Not regulatory advice. The binding text is each regulator's current official document; confirm the category with the authority before filing. FDA publishes no classification table: the US category follows FDA guidance or, where the guidance does not name the change, the closest example under 21 CFR 314.70.